Water Purification System
Water Purification System
- The Water Purification System (hereinafter referred to as the System) МКУБ.066136.001 is designed for producing water used in hydrogen generators, for preparing solutions, chromatographic samples, capillary electrophoresis, high-performance liquid chromatography (HPLC), and other analyzers that require ultrapure water. The System is connected directly to the municipal water supply line and produces high-purity deionized water with a specific electrical conductivity of 0.05 – 2.5 µS/cm. The instrument employs multi-stage water purification.
Technical Specifications
| Parameter | Value |
| Nominal deionized water output, l/h, not less than | 15 |
| Inlet water operating pressure, MPa | От 0,1 до 0,8 |
| Replacement interval for pre-filters, not more than, months (l) | 12(6000) |
| Replacement interval for reverse osmosis membrane, not more than, months (l) | От 18 до 24(3500) |
| Service life of ion-exchange filters*, not more than, l | 700) |
* Subject to compliance with the requirements for feed water according to SanPiN 2.1.3684-21.
- Mechanical filter removes dirt, suspended particles, insoluble impurities, silt, sand, colloidal compounds, rust, large inclusions (up to 10 microns), and other insoluble foreign matter from water;
- Carbon filters (two stages) absorb dissolved chlorine, other gases, organic compounds, and heavy metal salts from water. The service life of each carbon filter stage is 6,000 litres;
- Reverse osmosis membrane is a special membrane whose pores allow water molecules, oxygen, alcohols, methane derivatives, and trace organics to pass through, while retaining the main salts of light metals, bacteria, viruses, bacterial and viral metabolic products, and mould spores – with a selectivity of 96%;
- Ion-exchange filter block retains light metal salts that have passed through the membrane, reducing water conductivity to 0.06 – 0.1 µS/cm;
- All stages of the System are replaceable and can be easily exchanged for new ones. The specific electrical conductivity of the water at the system outlet is monitored by a built-in conductometer;
Submit application